OPERATION
61
2. Set the voltage range with the scroll wheel or
with the F1 ~ F4 soft-keys.
Range
AUTO, 600V(option), 310V, 155V
Soft-keys
AUTO, 600V(option), 310V, 155V
3. Press Enter to confirm the Range setting.
Range setting
Soft-key
settings
F1
F2
F3
F4
Note
If the range is changed from 155V to 600V, the
Irms and IPK values will automatically be changed
to a lower value. If the range is changed from 600V
to 155V, the Irms and IPK values remain the same.
If the voltage range is changed when the output is
on, the output will be automatically turned off.
Setting the Voltage Limit
Background
Setting the voltage limit allows the output
voltage to be set to any level within the voltage
limit (V Limit) range.
Steps
1. Press Shift + V to access the Volt
Limit menu.
Shift
+
V
V-Limit