3Dimensions System User Guide
Chapter 6: Images
MAN-05085-002 Revision 002 Page 81
DRAFT Preview Copy-Generated May 30, 2018
4. Acquire the images during the Optimal Imaging Period.
After you acquire an image, a marker will appear below the timer.
Figure 52: I-View 2D Contrast Screen, Optimal Imaging Period
5. Select the low and high exposure buttons to acquire both low and high energy
images.
Low: Conventional low-energy image
Sub: Contrasted, subtracted image
Low Sub: Shows both the low and sub screen images side-by-side in a split
screen.
Caution:
Using I-View 2D Contrast increases the recovery time the machine needs before more
images are acquired. When the x-ray tube status icon appears red
, a
recommended wait time is shown. This wait time lets the x-ray tube temperature
decrease to avoid damage that can void the x-ray tube warranty. Always make sure
that the x-ray tube status icon is green
before starting an I-View 2D Contrast
procedure. If the x-ray tube status icon turns red during a procedure, complete the
procedure quickly using the minimum number of exposures.