Fluent Pro Fluid Management System
English Introduction 5
Figure 2: Rear View of System
1.3.3.1 Fluid Bag Hooks
Fluid Bags (not included) containing hypotonic, isotonic, ionic and non-ionic distention fluids hang from the
Fluid Bag Hooks located on the IV pole at the top of the Fluent Pro Fluid Management System. Each Fluid Bag
Hook can support up to three (3) liters of fluid. Hang only a single Fluid Bag on the Fluid Bag Hooks at a time.
The In-FloPak
TM
contains a bag spike which connects the In-FloPak to the Fluid Bag. Fluid is pulled from the
Fluid Bag through the In-FloPak and delivered through the hysteroscope inflow channel into the patient.
WARNING!
Do not place excessive force or weight on the Fluid Bag Hooks. Doing so may result in an inaccurate fluid
deficit value, causing risk to patient safety.
WARNING!
When performing monopolar hysteroscopic electrosurgery, the distension medium must be electrically non-
conductive. Examples include glycine, sorbitol, and mannitol.
WARNING!
When performing bipolar electrosurgical resection procedures, the distension medium may be isotonic. Examples
include saline and lactated Ringer’s solution.
1.3.3.2 Touchscreen Monitor
The Touchscreen Monitor includes two system speakers and the Touchscreen User Interface. Tilt and swivel
the monitor for optimal viewing.
Caution!
Only use the Handle to move or position the Fluent Pro Fluid Management System. Do not pull or push the
system with the IV pole or Touchscreen Monitor.