Fluent Pro Fluid Management System
English Introduction 11
1.3.4 Fluent Pro Procedure Kit Components
Figure
9a: Carton Components
Figure
9b: Tray Components
Figure 9c: Waste Bag
WARNING!
Fluent Pro Procedure Kits are single use disposables. Do not re-use or reprocess any of the components contained within
the single-use Procedure Kit.
1.3.4.1 Fluent Pro In-FloPak
The In-FloPak pulls clean fluid from the Fluid Bag. It contains the Fluid Bag Tube and the hysteroscope Inflow
Tube. It fits securely into the blue Fluent Pro In-FloPak Receptacle on the front left side of the fluid
management system. These connections allow the transfer of distension fluid from a Fluid Bag to the
hysteroscope inflow channel. The flow of fluid is monitored and controlled using the Touchscreen to maintain
the pressure at a specified setting.
1.3.4.2 Fluent Pro Out-FloPak
The Out-FloPak drains waste fluid from the hysteroscope Outflow Channel, the MyoSure tissue removal device
(TRD) Tube, and the Under-Buttocks Drape (UBD) Tube into the Waste Bag. It fits securely into the yellow Fluent
Pro Out-FloPak Receptacle on the front right side of the fluid management system. These connections allow
the transfer of fluid from the hysteroscope Outflow Channel, the MyoSure tissue removal device (TRD), and the
Under-Buttocks Drape (UBD) Port to the Waste Bag.