Fluent Pro Fluid Management System
12 Introduction English
1.3.4.3 Tissue Trap
The Tissue Trap is designed to capture resected tissue throughout the procedure to allow the tissue to be sent
to pathology. The Tissue Trap Canister contains the Tissue Trap that captures the resected tissue. Please
monitor the Tissue Trap so it does not overfill. For details on handling the Tissue Trap, see Section 3.8.3.
1.3.4.4 Waste Bag
The non-sterile Waste Bag is designed to capture waste fluid from hysteroscopic procedures. The Waste Bag
hangs on the Waste Bag Hooks at the bottom of the Fluent Pro Fluid Management System. Hang only a single
Waste Bag on the Waste Bag Hooks at a time. Hanging more than one bag may impact fluid deficit accuracy.
The Waste Bag includes an attached cap. For details on replacing the Waste Bag, see Section 3.8.2.
If an accurate manual deficit assessment is required, pour fluid into calibrated container. The markings on the
Waste Bag are not intended as a measuring device, are only for general reference and are not intended as a
specific volumetric measuring device.