Fluent Pro Fluid Management System
English During Procedure 41
3.6.5 The Fluent Pro Out-FloPak will continue to evacuate fluid to ensure an accurate deficit value and all resected tissue is
captured.
WARNING!
Do not remove the Fluent Pro Out-FloPak and Waste Bag if you want to allow suction to continue to run to remove excess
fluid from the Under-Buttocks Drape (UBD) to accurately reflect the deficit. If the Fluid Bag or Waste Bag are removed at
this time, the deficit will need to be calculated manually.
3.6.6 When you feel the excess fluid has been removed from the Under-Buttocks Drape (UBD), touch the ‘DONE’ icon.
3.6.7 The system displays a Procedure Complete screen that displays the following final procedure data:
• Fluid Deficit
• Total Fluid In
• Cutting Time
3.6.8
To exit the Procedure Complete screen, perform the following steps:
3.6.8.1 Retrieve Tissue Trap (specimen).
3.6.8.2 Remove disposables.
3.6.9 Once FloPaks have been removed from the system, the ‘EXIT’ icon will be activated.
3.6.10 Select ‘EXIT’. The Launch screen will be displayed, allowing you to begin a new procedure or to access the last procedure
data.
3.6.11 For specifics on disassembly and disposal see Section 4.1.
Note: If you forget to record the results at the end of the procedure, the Fluent Pro Fluid Management System will retain the
results from the last procedure. See below for details on how to view last procedure data.
3.7
Last Procedure Data
3.7.1 To access data from the last procedure, prior to starting a new procedure, perform the following steps:
3.7.1.1 On the System Setup screen, touch the ‘LAST PROCEDURE DATA’ icon.
3.7.1.2 The following final data will be displayed from the last procedure:
• Fluid Deficit
• Total Fluid In
• Cutting Time