Fluent Pro Fluid Management System
English Supplementary Information 77
Emissions test Compliance Electromagnetic environment - guidance
RF emissions CISPR 11 Class A
The Fluent Pro Fluid Management System is suitable for use in all establishments
other than domestic and those directly connected to the public low-volt- age power
supply network that supplies buildings used for domestic purposes.
Note: The EMISSIONS characteristics of this equipment make it suitable for use in
industrial areas and hospitals (CISPR 11 class A). If it is used in a residential
environment (for which CISPR 11 class B is normally required) this equipment might
not offer adequate protection to radio-frequency communication services. The user
might need to take mitigation measures, such as relocating or re-orienting the
equipment.
Harmonic emissions
IEC 61000-3-2
Complies
Voltage fluctuations / flicker
emissions IEC 61000-3-3
Complies
6.4.2 Electromagnetic Immunity Guidance and Manufacturer’s Declaration
The Fluent Pro Fluid Management System is intended for use in the electromagnetic environment specified below. The
customer or the user of the Fluent Pro Fluid Management System should ensure that it is used in such an environment.
Non-observance of the instructions listed in this manual can lead to damage or malfunction of the system.
Table 6: Electromagnetic Immunity Guidance
Immunity test
IEC 60601
test level
Compliance level Electromagnetic environment guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
± 8 kV contact
± 2, ± 4, ± 8, ± 15 kV
air
± 8 kV contact
± 2, ± 4, ± 8, ± 15 kV
air
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30%.
Electrical fast transients / bursts
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input /
output lines
± 2 kV for power
supply lines
± 1 kV for input/output
lines
Mains power quality should be that of a typical
commercial or hospital environment.
Surge IEC 61000-4-5 ± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips, short interruptions,
and voltage variations on power
supply input lines IEC 61000-4-11
0% UT (100% dip in
the UT) for ½ cycle at
0°, 45°, 90°, 135°,
180°, 225°, 270°, and
315°
0% UT (100% dip in
the UT) for ½ cycle at
0°, 45°, 90°, 135°,
180°, 225°, 270°, and
315°
Mains power quality should be that of a typical
commercial or hospital environment.
If the user/operator of the Fluent Pro Fluid
Management System requires continued
operation during power mains interruptions, it is
recommended that the Fluent Pro Fluid
Management System be powered from an
uninterruptible power supply or battery.
0% UT (100% dip in
the UT) for 1 cycle and
70% (30% dip in the
UT) UT for 25/30 cycles
at 0°
0% UT (100% dip in
the UT) for 1 cycle and
70% (30% dip in the
UT) UT for 25/30 cycles
at 0°
0% UT (100% dip in
the UT) for 250/300
cycles
0% UT (100% dip in
the UT) for 250/300
cycles
Power frequency (50/60 Hz)
magnetic field IEC 61000-4-8
30 A/m 30 A/m
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.