EasyManua.ls Logo

Hologic Fluent Pro - Priming the System; Figure 32: Initial Prime Screen; Figure 33: Unclamp Fluid Bag

Hologic Fluent Pro
98 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Fluent Pro Fluid Management System
26 Procedure Setup English
2.7.6.11 Touch ‘NEXT on the Touchscreen.
2.7.6.12 The Initial Prime screen is then displayed.
Figure 32: Initial Prime Screen
2.7.7 Priming the system
2.7.7.1 Begin each procedure with a new, full Fluid Bag. Priming the system runs the pump for approximately one (1)
minute to purge air from the tubing and calibrate the hysteroscope.
2.7.7.2 Before priming, make sure that the Fluent Pro In-FloPak and Fluent Pro Out-FloPak are properly connected as
described in Section 2.7: Setup System.
2.7.7.3 The system should be primed at the beginning of a case. Refer to Section 2.7.9 for instructions on how to Reprime
the system.
2.7.7.4 To prime the system, perform the following steps:
2.7.7.4.1 Unclamp Fluid Bag.
Figure 33: Unclamp Fluid Bag

Table of Contents

Related product manuals