Fluent Pro Fluid Management System
62 Troubleshooting English
WARNING!
When replacing a fuse, use only type T5AH, 250V fuses.
5.3.8.2.1 The Power Cord is properly connected to both the Power Port on the rear of the Fluent Pro Fluid
Management System and to a wall outlet.
5.3.8.2.2 The wall outlet has power. Test the outlet by plugging in another device to ensure it is working
properly.
5.3.8.2.3 If the system has adequate power running to it but it will still not power on, the fuse may be
defective. To replace the fuse:
5.3.8.2.4 Disconnect the Power Cord from the wall outlet.
5.3.8.2.5 Obtain the appropriate fuse(s) for replacement. The approved fuse type is T5AH, 250 V. There are
2 per system.
5.3.8.2.6 Locate the fuse holder located in the power module on the rear of the system. Using a small
flathead screwdriver, remove the fuse holder.
5.3.8.2.7 Pull the fuse out of the fuse holder and inspect for damage.
5.3.8.2.8 Insert the new fuse(s). Use only the type of fuse: T5AH, 250 V fuses.
5.3.8.2.9 Insert the fuse holder.
5.3.8.2.10 Reconnect the Power Cord to the wall outlet power then power on the system and follow on-
screen prompts. To resume the previous procedure, refer to Section 5.2 and turn the system on to
ensure it is working properly.
5.3.9
System Frozen
5.3.9.1 If the system is frozen and cannot be shut down by touching the Power Button, perform a hard shutdown by
holding the Power Button down for 5 seconds. Power on the system and follow on-screen prompts. To resume
the previous procedure, refer to Section 5.2.