Fluent Pro Fluid Management System
76 Supplementary Information English
Item Specification
Pressure (mmHg) ± 35
Flow (mL/min) ± 50
Deficit (mL) ± 50 under normal use
Dimensions (Width x Height x Depth) 23in x 67in x 27in / 584mm x 1702 mm x 686mm
System Weight 65Kg
Console Mass 45Kg
Safe Working Mass 20 Kg
Interfaces
Signal IN/OUT components None
Mains connection IEC 60320-1 C14
Maximum Load Cell Capacity <6.3Kg for Supply and Waste Bag Hooks
System Useful Life The system shall have a useful life up to 5 years.
6.3 Power Cord Safety
Ensure the connection data and technical specifications of the power supply comply with DIN VDE or national requirements. The wall
outlet power supply cord must be plugged into a properly installed safety wall plug (see DIN VDE 0107). Read the device label located
in rear of pump to determine the operating voltage of the system.
The power connection must be equipped with a grounding contact. Use the Fluent Pro Fluid Management System power cord to
establish a connection between the wall outlet and the power cord connection located in the rear of the system.
Only for U.S. operators: Use only a certified (UL-listed), removable power cord, type SJT, minimal 16 AWG, 3 leads. The plug
connectors must comply with NEMA 5-15 or IEC 320/CEE22. Grounding will only be reliable if the equipment is connected to a
corresponding hospital grade outlet.
Integrate the system into the potential equalization system as specified by local safety rules and regulations.
Medical devices are subject to special safety and protective measures concerning electromagnetic compatibility (hereafter
abbreviated as EMC).
This system is to be used only for the purposes described in the manual and must be installed, set up, and operated in compliance
with the EMC notes and instructions. See Section 6.4: Electromagnetic Compatibility.
6.4
Electromagnetic compatibility
6.4.1 Electromagnetic Emissions Guidance and Manufacturer’s Declaration
The Fluent Pro Fluid Management System is intended for use in the electromagnetic environment specified below. The
customer or the user of the Fluent Pro Fluid Management System should ensure that it is used in such an environment.
Non-observance of the instructions listed in this manual can lead to damage or malfunction of the system.
WARNING!
Use of accessories, transducers, and cables other than those specified or provided by the manufacturer of this equipment
could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result
in improper operation.
Table 5: Electromagnetic Emissions Guidance
Emissions test Compliance Electromagnetic environment - guidance
RF emissions CISPR 11 Group 1
The Fluent Pro Fluid Management System uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are not likely to cause
interference in nearby electronic equipment.
RF emissions CISPR 11 Group 2
The Fluent Pro Fluid Management System must emit electro-magnetic energy in
order to perform its intended function. Nearby electronic equipment may be
affected.