Fluent Pro Fluid Management System
64 Troubleshooting English
5.3.10.2 Alerts and Troubleshooting (screen flashing Red and Pink)
Table 2: Real-time Display, Alerts and Troubleshooting
Alert Message Description Resolution
Waste Bag Full Volume in Waste Bag has reached 6000mL
Remove full Waste Bag and replace with a new empty
Waste Bag.
Waste Bag Missing Weight on Waste Bag Hooks is less than 90g
Hang a Waste Bag on the Waste Bag Hooks. Ensure
the bag is level on both Hooks.
Fluid Bag(s) Missing Weight on both Fluid Bag Hooks is less than 50g each Hang a Fluid Bag on either Fluid Bag Hook
Fluid Bag(s) Empty Weight on Fluid Bag Hooks is less than 175g
Replace empty Fluid Bag(s) with new Fluid Bag(s).
Press clear on the GUI.
Excessive Movement of
Fluid Bag(s) or Nearly
Empty Fluid Bag(s)
Fluid Bag Hooks are unstable for at least 25 seconds
Stabilize Fluid Bag(s) or Replace Fluid Bag(s). Do not
touch the Fluid Bag Hooks or move the system. Press
clear on the GUI.
Excessive Movement of
Waste Bag
Waste Bag Hooks are unstable for at least 5 seconds
Stabilize Waste Bag. Do not touch the Waste Bag
Hooks or move the system. Press clear on the GUI.
Fluid Bag(s) Clamped or
Not Spiked
Fluid is not flowing through the Fluid Bag tubing to the
system.
Check that Fluid Bag(s) are spiked and that tubing is
unclamped. Press clear on the GUI.
Deficit Limit
Reached/Exceeded
The actual deficit is greater than or equal to the set
deficit limit
Consider increasing the deficit limit, if clinically
appropriate.
Front Fluid Bag Hook
Overloaded
Excessive weight is detected on the Front Fluid Bag
Hook
Do not lean on the Fluid Bag Hooks. Remove any
inappropriate items from the hook. Remove excessive
full Fluid Bags from the hook. Press clear on the GUI.
Back Fluid Bag Hook
Overloaded
Excessive weight is detected on the Front Fluid Bag
Hook
Do not lean on the Fluid Bag Hooks. Remove any
inappropriate items from the hook. Remove excessive
full Fluid Bags from the hook. Press clear on the GUI.
Waste Bag Hooks
Overloaded
Excessive weight is detected on the Front Fluid Bag
Hook
Do not lean on the Waste Bag Hooks. Remove any
inappropriate items from the hook. Ensure only one (1)
Waste Bag is hanging and remove any extra bags.
Press clear on the GUI.
In-FloPak Error
The lock pin did not engage with the In-FloPak. In-
FloPak is not locked.
Depress the release lever and remove the In-FloPak.
Re-install the In-FloPak.
In-FloPak Not Found In-FloPak was removed or is not fully inserted.
Depress the release lever and remove the In-FloPak.
Reinstall the In-FloPak. Ensure the In-FloPak is flush
with the Console.
Out-FloPak Not Found Out-FloPak was removed or is not fully inserted.
Depress the release lever and remove the Out-FloPak.
Reinstall the Out-FloPak. Ensure the Out-FloPak is
flush with the Console.
Tubing Pressure Too High
The inflow tubing or hysteroscope is blocked in a way
that is causing high pressure in the inflow tube.
Open Inflow stopcock or unclamp tubing. Press clear
on the GUI.
Priming Error- Flow
Obstructed
Inflow stopcock is not fully open or the end of the
hysteroscope is obstructed
Open the inflow stopcock fully and ensure there is
nothing obstructing the tip of the hysteroscope. Press
clear on the GUI and reprime the system.