Fluent Pro Fluid Management System
52 After Procedure English
4.1.6.3 Depress the In-FloPak lever and remove the In-FloPak from the Fluent Pro Fluid Management System console.
4.1.6.4 Dispose of the Fluent Pro In-FloPak according to your facility’s protocol.
4.1.7
Disassemble the Fluent Pro Out-FloPak
4.1.7.1 Depress the Out-FloPak lever and remove the Out-FloPak from the Fluent Pro Fluid Management System
console.
4.1.7.2 Dispose of the Fluent Pro Out-FloPak according to your facility’s protocol.
4.1.8
Shutdown and Disassemble the Fluent Pro Fluid Management System
4.1.8.1 Push the Power Button to power down the system.
4.1.8.2 The system displays a message confirming that you want to end the procedure. The system will ding while the
confirmation screen is active.
4.1.8.3 To power down the system, touch the ‘YES icon.
4.1.8.4 A powering down screen will appear along with a status bar. Do not unplug the system until the sequence is
complete and the screen is dark. When the system is powered down and plugged in, the power button will be
illuminated orange.
4.1.8.5 Store the foot pedal in the Foot Pedal Basket.
4.1.8.6 Unplug the Fluent Pro Fluid Management System console Power Cord.
4.1.8.7 Wrap the Power Cord around the cord wrap on the back of the system.
4.2
Cleaning the Fluent Pro Fluid Management System
4.2.1 Ensure the Fluent Pro Fluid Management System is disconnected from the electrical source.
4.2.2 Wipe the surface of the system with a soft cloth moistened with disinfectant. For information regarding concentration of
disinfectant, refer to documentation provided by the manufacturer of the disinfectant. Examples of disinfectant include:
5% dishwashing soap in water
0.14% Ammonium chloride (PDI Sani-Cloth AF3, PDI Sani-Cloth HB, or equivalent)
Commercially available isopropyl alcohol solution (typically 70% isopropyl alcohol by volume, undiluted)
10% chlorine bleach and water solution (consisting of one part commercially available bleach and nine parts water)
Caution!
Do not sterilize or immerse the Fluent Pro Fluid Management System in disinfectant.
4.3
Storing the System
4.3.1 Prior to storing the system, complete Section 4.1 and 4.2.
4.3.2 To unlock the wheels, use your foot to lift (upward) the lever on each Wheel Lock.
4.3.3 Transport the system using the 360° Handle.
Caution!
Only use the Handle to move or position the Fluent Pro Fluid Management System. Do not pull or push the system with the
IV pole or Touchscreen Monitor.
WARNING!
Do not place excessive force or weight on the Fluid Bag Hooks. Doing so may result in an inaccurate fluid deficit value,