Fluent Pro Fluid Management System
42 During Procedure English
Figure 52: Last Procedure Data Screen
Note: In the event of a power failure, refer to Section 5.2 for retrieval of last procedure data.
3.8
Replacing Components
Depending on the procedure, you may need to replace disposable components. This Section provides information and instructions for
replacing the Fluid Bag, Waste Bag, and Tissue Trap while using the Fluent Pro System. To alert you of changes, the Fluent Pro Fluid
Management System Fluid Bag Hooks and Waste Bag Hooks contain built-in scales, and the system will prompt you if you need to
take action. The System will not alert you if the tissue trap is near capacity or full, please monitor tissue trap for capacity.
3.8.1
Adding the Fluid Bag
3.8.1.1 The system does not need to be paused in order to add or change a Fluid Bag, as long as a second bag is
hung and unclamped.
3.8.1.2 Replace the Fluid Bag when the Fluid Bag is almost empty or when alerted by the system.
3.8.1.3 If leaving an empty bag on Hooks, ensure that empty Fluid Bag is clamped.
3.8.1.4 Hang only a single Fluid Bag on the Fluid Bag Hooks at a time.
3.8.1.5 A minimum of one Fluid Bag should always be unclamped when the system is running.
3.8.1.6 Perform the following steps to replace the Fluid Bag during a procedure:
WARNING!
Do not place excessive force or weight on the Fluid Bag Hooks. Doing so may result in an inaccurate fluid
deficit value, causing risk to patient safety.
Only hang one (1) Fluid bag per supply hook. Hanging more than 1 fluid bag per supply hook can affect the
system’s ability to recognize empty fluid bags during procedure.
3.8.1.6.1 One Fluid Bag on Fluid Bag Hooks
3.8.1.6.1.1 Hang a new Fluid Bag on the vacant hook.
3.8.1.6.1.2 The total volume hung on each Fluid Bag Hook should not exceed 3L of fluid.
3.8.1.6.1.3 Spike the new Fluid Bag.
3.8.1.6.1.4 Open clamp on the new Fluid Bag.
3.8.1.6.1.5 Close the clamp on the empty Fluid Bag.
3.8.1.6.2 Two Fluid Bags on Fluid Bag Hooks (One Bag Unclamped)