Fluent Pro Fluid Management System
English During Procedure 39
3.5.6.2 Touch the CHANGE SCOPE icon. The Ready to Prime screen will appear.
Figure 48: Ready to Prime Screen
3.5.6.3 Direct reprime fluid the Under-Buttocks Drape (UBD).
3.5.6.4 Before beginning reprime, ensure Fluid Bag has at least 200mL of fluid to complete the reprime process. If
there is not enough fluid, replace the Fluid Bag with a new, full Fluid Bag.
3.5.6.5 Prime should be performed at the operative height.
WARNING!
Do not prime inside the patient.
3.5.6.6 Follow the prompts and touch the ‘PRIME icon.
3.5.6.7 When the prime and calibration of the hysteroscope are complete, the Priming Complete screen will be
displayed.
Figure 49: Priming Complete Screen