6.2 Technical Specifications
Table 4: Technical Specifications
Item Specification
Model or type designation FLT-200
Mains voltage range [V] 100-240 VAC
Supply Frequency Range [Hz] 50-60Hz
Fuse designation T5AH, 250 V fuses
Mode of Operation Non-continuous (Max Duty: 60 minutes ON, 30 minutes IDLE)
Power Consumption 350W
Upper Voltage Range
Normal Operation 240 VAC
Peak 264 VAC
Lower Voltage Range
Normal Operation 120 VAC
Peak 90 VAC
Protection class (I, II, III) I
Application part type (B, BF, CF) BF
Defibrillator protected (yes, no) No
Protection type (IP code) IP21
Conformity with the following standards IEC 60601-1:2005 + A1:2012 + A2:2020
IEC 60601-1-6:2010 + A1:2013 + A2:2020
IEC 62366-1:2015 + A1:2020
IEC 62304:2006 +A1:2015
IEC 60601-1-2:2014 + A1:2020
Operating conditions 14-30ºC / 57-86ºF
20-80% rel. humidity, Non-Condensing
8000 ft (2438.4 Meters) max. altitude above sea level for use
Atmospheric pressure: 75-101 kPa
Storage and transportation conditions -30-60ºC / -22-140ºF
15-90% rel. humidity, Non-Condensing
Use possible with flammable anesthetic
gases
No
Maximum sound level ≤75 decibels (dbA at 1meter)
Adjustable Values
Pressure range (mmHg) 40 - 150
Deficit Limit (mL) 100 - 2500 at start of procedure
100 - 9950 after procedure has started
Measurement range
Flow (mL/min) 0-1000
Deficit (mL) ±9999
Accuracy