ThinPrep™ 5000 Processor with AutoLoader Operator’s Manual
8.13
MAINTENANCE
WARNING:
Hazardous Waste
Toxic Mixture
Flammable Liquid and Vapor
7. Place the transport cover onto the waste bottle for transporting to the waste disposal area.
8. Dispose of the liquid waste from the waste bottle according to your laboratory guidelines.
9. Before reattachment, inspect the O-ring seal on the inside of the waste cap for debris.
See Figure 8-14.
• If debris is present, clean the seal with water using a lint-free wipe.
• Apply a thin layer of vacuum grease to the O-ring.
Figure 8-14 Inspect Waste Bottle O-Ring Seal
10. Return the waste bottle back to its original location and retighten the waste cap onto the
bottle.
• Verify that the waste cap is firmly tightened and confirm that the waste tubing is not pinched
or twisted.
Press the
Next
button to perform a leak test. This repressurizes the waste bottle and checks that the
system can hold pressure. It also measures the fluid level to verify that the waste bottle has been
emptied. See Figure 8-15.
Note:
The leak test MUST be run after emptying the bottle.