INFICON Connecting to a Sample | 7
074-594-P1H Micro GC Fusion Operating Manual 141 / 319
7.4 Sampling Using Gas Cylinders and Canisters
Gas cylinders and canisters can be connected directly to the 1/16 in. Swagelok or 1/8
in. Quick Connect sample inlet. Sample pressures must not exceed 172 kPa (25 psi),
or 6,895 kPa (1,000 psi) when the optional Integrated Sample Conditioner is installed.
CAUTION
Sample inlet gas pressure must be less than 172 kPa (25 psi). Exceeding 172
kPa (25 psi) will damage Micro GC Fusion and require factory repair.
CAUTION
When using the Integrated Sample Conditioner, sample inlet pressure must
not exceed 6,895 kPa (1,000 psi).
Exceeding 6,895 kPa (1,000 psi) will damage Micro GC Fusion and require factory
repair.