INFICON Checkout Gas Chromatograms | 14
074-594-P1H Micro GC Fusion Operating Manual 315 / 319
14.3.2 Natural Gas Checkout Gas Using Helium Carrier Gas
Peak Number Compound Concentration (mole %)
1 nitrogen 1.50
2 methane 75.32
3 carbon dioxide 1.50
4 ethane 8.99
5 propane 6.00
6 isobutane 3.01
7 n-butane 2.00
8 isopentane 1.02
9 n-pentane 0.50
10 n-hexane 0.10
11 n-heptane 0.05
1
2
3
4
5
6
7
8
4
9
10
11
•
Column Temperature Profile
: 55°C (20 s hold) -> 230°C (50 s hold), 1°C/s
•
Column Pressure:
30 psi
•
Inject (ms) Time:
30 ms