INFICON Connecting to a Sample | 7
074-594-P1H Micro GC Fusion Operating Manual 143 / 319
7.5.2 Installing a Syringe Luer Lock Adapter onto a Gas Tight
Syringe
1
Attach the luer fitting of the Syringe Luer Adapter (PN 952-024-G1) to the PTFE
luer lock termination of the gas tight syringe.
2
Turn the luer fitting on the Syringe Luer Adapter clockwise until finger tight.
7.6 Sampling Using Bags
1
Open a new sample bag.
2
Connect the sample bag to the sample source.
3
Open the sample source.
4
Fill the sample bag, close it, and disconnect from the sample source.
5
Close the sample source.
6
Connect the sample bag to the Micro GC Fusion inlet using stainless steel or
Teflon tubing.
An adapter may be required to connect the tubing to the Micro GC Fusion 1/16
in. Swagelok front inlet.
7
Open the sample bag when the GC is ready to sample.