EasyManua.ls Logo

Inficon Micro GC Fusion 2-Module

Inficon Micro GC Fusion 2-Module
319 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
INFICON Software Operation | 9
074-594-P1H Micro GC Fusion Operating Manual 171 / 319
9.4.11 Edit Sequence
1
In the
Methods/Sequences
window, select the desired sequence.
2
Click the icon to edit the highlighted sequence. The
Sequence Editor
is
displayed.
3
Click the step shortcut menu to select an action. Available actions include:
ð
Load Method
- Loads the desired method
ð
Run Method
- Runs the method selected in Load Method, once the method
set points have been reached.
ð NOTE: Separate
Load Method
and
Run Method
commands allow the user to
input a delay between the two.
ð
Load and Run Method
- Loads the selected method. Once the method set
points are reached, a run is initiated
ð
Calibration Runs
- Loads and runs the selected method. Once all runs are
complete, the area of each compound will be averaged and replaced in the
desired calibration level. See Auto-Calibration [}228].
4
Click the method shortcut menu to select a method.

Table of Contents

Related product manuals