INFICON Software Operation | 9
074-594-P1H Micro GC Fusion Operating Manual 231 / 319
9.10.11 Reprocess Runs
Often it is desirable to edit integration parameters and adjust retention time windows
after a run is completed, without adjusting the calibration information.
After a run is performed:
1
In the data browser, click on the run.
2
Click the icon to open the calibration window. By default,
Module A
will
display.
3
Click the icon to select another module, if desired.
4
Integration parameters can be edited in the
Integration
tab. The retention time
windows can be edited in the
Compounds
tab.