INFICON Routine Operation | 10
074-594-P1H Micro GC Fusion Operating Manual 283 / 319
6
Use forceps to remove the used filter from the unit.
7
Use forceps to place a new filter into the unit. Ensure the flat side of the filter
faces the forceps.
8
Replace the hex plug.
9
Reconnect the external electrical cables, including the power cable and Ethernet
cable, sample gas connection(s), carrier gas connection(s) and venting
connections.
10
Micro GC Fusion is now ready to use.