EasyManua.ls Logo

Inficon Micro GC Fusion 2-Module

Inficon Micro GC Fusion 2-Module
319 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
INFICON Troubleshooting | 11
074-594-P1H Micro GC Fusion Operating Manual 285 / 319
Symptom Possible sources of error Remedy
There is an internal leak. Contact INFICON. See Service and
Technical Support [}299].
There is a malfunction in the
column or injector.
Check the flow rate of the columns by
attaching a flow meter to the analytical and
reference vents on the back of Micro GC
Fusion
If there is no flow from either or both vents,
contact INFICON. See Service and
Technical Support [}299].
There is a malfunction in the
detector.
Contact INFICON. See Service and
Technical Support [}299].
The chromatogram is
displaying a flat line with
no peaks
The
Inject (ms)
time method
parameter is set to 0 ms.
Change the
Inject (ms)
time to a value other
than 0 ms.
The
Sample Pump Time
method
parameter is set to 0 seconds.
Ensure that the
Sample Pump Time
is set for
at least 15 seconds.
There is a malfunction in the
column or injector.
Check the flow rate of the columns by
attaching a flow meter to the analytical and
reference vents on the back of Micro GC
Fusion
If there is no flow from either or both vents,
contact INFICON. See Service and
Technical Support [
}299].
There is a malfunction in the
sample pump or the sample
pump does not turn on.
Listen for the noise from the pump when
taking a sample. If there is no noise, contact
INFICON.
There is a malfunction in the
detector.
Contact INFICON. See Service and
Technical Support [}299].
A peak is missing The
Backflush Time
method
parameter is set too early (if
applicable).
Increase the
Backflush Time
until all peaks
of interest appear on the chromatogram.
A peak is saturated The
Inject (ms)
time method
parameter is set too long.
Shorten the
Inject (ms)
time.
The
Injector Temperature
method parameter is set too
high.
Lower the
Injector Temperature.
Chromatograms are not
appearing/no data is
showing
The web browser requires a
restart or refresh.
Restart or refresh the web browser.

Table of Contents

Related product manuals