INFICON Installation | 6
074-594-P1H Micro GC Fusion Operating Manual 83 / 319
12
Tighten the Valco fitting on the gas-liquid separator using a 1/4 in. wrench.
13
Use additional tubing to connect the gas-liquid separator sample vent to a fume
hood or other safe location.
14
Reconnect all electrical cables, including the power cable and Ethernet cable.
15
Reconnect the carrier gas and venting connections.
16
Turn on Micro GC Fusion. Connect a sample to the sample inlet on the gas-liquid
separator to perform an analysis.
WARNING
Safely vent carrier and sample streams - potentially toxic, noxious, asphyxiant
(oxygen displacing) or flammable gases - outside Micro GC Fusion and away
from the operating area. Vent toxic gases to a fume hood, chemical trap, or
reaction medium.
CAUTION
Ensure the vent tubing is not inclined upward to prevent potential liquid
accumulation in the tubing.
CAUTION
Blocking the vent port during operation significantly increases the output
pressure from the device. When this happens, the instrument may be
damaged if the incoming sample pressure is greater than 172 kPa (25 psi).