:TEMPerature
:NPLCycles <n>
:NPLCycles?
:REFerence <n>
:STATe <b>
:STATe?
:ACQuire
:REFerence?
:DIGits <n>
:DIGits?
:AVERage
:TCONtrol <name>
:TCONtrol?
:COUNt <n>
:COUNt?
:STATe <b>
:STATe?
:TCouple
:TYPE <name>
:TYPE?
:RJUNction
:RSELect <name>
:RSELect?
:SIMulated <n>
:SIMulated?
:REAL
:TCOefficient <n>
:TCOefficient?
:OFFSET <n>
:OFFSet?
Path to configure temperature:
Set integration rate (line cycles; 0.01 to 10).
Query line cycle integration rate.
Specify reference; -200 to 1372
Enable or disable reference.
Query state of reference.
Use input signal as reference.
Query reference value.
Specify measurement resolution (4 to 7).
Query resolution.
Path to configure and control the filter.
Select filter type: (MOVing or REPeat).
Query filter type.
Specify filter count (1 to 100).
Query filter count.
Enable or disable filter.
Query state of digital filter.
Path to configure thermocouple:
Select thermocouple type (J, K, or T).
Query thermocouple type.
Path to configure reference junction:
Select reference type (SIMulated or REAL).
Query reference type.
Specify simulated temperature in °C (0 to 50).
Query simulated temperature.
Path to configure real reference junction:
Specify temp coefficient (-0.09999 to 0.09999).
Query temp coefficient.
Specify voltage offset at 0°C (-0.09999 to 0.09999).
Query voltage offset.
1
0
OFF
6
(Note)
10
OFF
J
SIMulated
23°C
2e-4
5.463e-2
:FREQuency
:APERture
:APERture?
:THReshold
:VOLTage
:RANGe <n>
:RANGe?
:REFerence <n>
:STATe <b>
:STATe?
:ACQuire
:REFerence?
:DIGits <n>
:DIGits?
Path to configure frequency.
Sets gate time for period measurements (0.01-1.0s).
Query period gate time.
Path to select the threshold voltage range:
Select threshold range (0 to 1010).
Query threshold range.
Specify reference (0 to 1.5e7)
Enable or disable reference.
Query state of reference.
Use input signal as reference.
Query reference value.
Specify measurement resoltuion (4 to 7).
Query resolution.
10
0
OFF
7
Table 5-6(cont.)
SENSe command summary
Command Description
Default
parameter
SCPI
5-14 SCPI Command Reference

Table of Contents

Other manuals for Keithley 2000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2000 and is the answer not in the manual?

Keithley 2000 Specifications

General IconGeneral
Digits6.5
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Resistance Range100 Ω to 100 MΩ
ConnectivityIEEE-488 (GPIB)
Display TypeVFD