EasyManua.ls Logo

Keithley 2002

Keithley 2002
366 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
IEEE-488 Reference
3-66
Table 3-16
UNIT command summary
Command Description
Default
parameter SCPI
:UNIT
:TEMPerature <name>
:TEMPerature?
:VOLTage
:AC <name>
:DB
:REFerence <n>
:REFerence?
:DBM
:IMPedance <n>
:IMPedance?
:AC?
Select temperature measurement units (C, CEL, F, FAR or K).
Query temperature units.
Path to configure ACV units.
Select ACV measurement units (V, DB or DBM).
Path to set DB reference voltage.
Specify reference in volts.
Query DB reference.
Path to set DBM reference impedance.
Specify reference impedance (1-9999).
Query DBM reference impedance.
Query ACV units.
C
V
1
75