IEEE-488 Reference
3-105
PRINT #1, "output 16; :volt:dc:aver:adv:ntol 20; ntol?" ' Get noise
tolerance
PRINT #1, "enter 16" ' Get response from
2002
[:STATe] <b>
[:SENSe[1]]:CURRent:AC:AVERage:ADVanced[:STATe] <b> Control advanced filter for ACI
[:SENSe[1]]:CURRent:DC:AVERage:ADVanced[:STATe] <b> Control advanced filter for DCI
[:SENSe[1]]:VOLTage:AC:AVERage:ADVanced[:STATe] <b> Control advanced filter for ACV
[:SENSe[1]]:VOLTage:DC:AVERage:ADVanced[:STATe] <b> Control advanced filter for DCV
[:SENSe[1]]:RESistance:AVERage:ADVanced[:STATe] <b> Control advanced filter for Ω2
[:SENSe[1]]:FRESistance:AVERage:ADVanced[:STATe] <b> Control advanced filter for Ω4
<b> = 0 or OFF Disable advanced filter
1 or ON Enable advanced filter
[:STATe]? Query state of advanced filter
These commands are used to enable or disable the advanced filter for the specified function. The
:ADVanced:NTOLerance commands are used to set tolerance levels.
PRINT #1, "output 16; :volt:dc:aver:adv on; adv?" ' Enable advanced filter
PRINT #1, "enter 16" ' Get response from 2002
:DETector commands
[:FUNCTION] <name>
[:SENSe[1]]:CURRent:AC:DETector[:FUNCTION] <name> Specify detector for ACI
[:SENSe[1]]:VOLTage:AC:DETector[:FUNCTION] <name> Specify detector for ACV
ACI:
<name> = RMS AC RMS amps measurements
AVERage AC average amps measurements
ACV:
<name> = RMS AC RMS volts measurements
AVERage AC average volts measurements
LFRMs Low frequency AC volts measurements
PEAK AC peak volts measurements
NPEak AC negative peak spike volts measurements
PPEak AC positive peak spike volts measurements
[:FUNCtion]? Query selected measurement type
These commands are used to select the measurement type for the ACI and ACV functions. Refer
to paragraphs 2.4.1 (ACV) and 2.4.2 (ACI) for details on AC-TYPE.
PRINT #1, "output 16; :curr:ac:det aver; det?" ' Select average detector
PRINT #1, "enter 16" ' Get response from 2002
Program fragment
Parameters
Query
Description
Program fragment
3.18.12
Parameters
Query
Description
Program fragment