EasyManuals Logo

Keithley 2002 User Manual

Keithley 2002
366 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #228 background imageLoading...
Page #228 background image
IEEE-488 Reference
3-105
PRINT #1, "output 16; :volt:dc:aver:adv:ntol 20; ntol?" ' Get noise
tolerance
PRINT #1, "enter 16" ' Get response from
2002
[:STATe] <b>
[:SENSe[1]]:CURRent:AC:AVERage:ADVanced[:STATe] <b> Control advanced filter for ACI
[:SENSe[1]]:CURRent:DC:AVERage:ADVanced[:STATe] <b> Control advanced filter for DCI
[:SENSe[1]]:VOLTage:AC:AVERage:ADVanced[:STATe] <b> Control advanced filter for ACV
[:SENSe[1]]:VOLTage:DC:AVERage:ADVanced[:STATe] <b> Control advanced filter for DCV
[:SENSe[1]]:RESistance:AVERage:ADVanced[:STATe] <b> Control advanced filter for 2
[:SENSe[1]]:FRESistance:AVERage:ADVanced[:STATe] <b> Control advanced filter for 4
<b> = 0 or OFF Disable advanced filter
1 or ON Enable advanced filter
[:STATe]? Query state of advanced filter
These commands are used to enable or disable the advanced filter for the specified function. The
:ADVanced:NTOLerance commands are used to set tolerance levels.
PRINT #1, "output 16; :volt:dc:aver:adv on; adv?" ' Enable advanced filter
PRINT #1, "enter 16" ' Get response from 2002
:DETector commands
[:FUNCTION] <name>
[:SENSe[1]]:CURRent:AC:DETector[:FUNCTION] <name> Specify detector for ACI
[:SENSe[1]]:VOLTage:AC:DETector[:FUNCTION] <name> Specify detector for ACV
ACI:
<name> = RMS AC RMS amps measurements
AVERage AC average amps measurements
ACV:
<name> = RMS AC RMS volts measurements
AVERage AC average volts measurements
LFRMs Low frequency AC volts measurements
PEAK AC peak volts measurements
NPEak AC negative peak spike volts measurements
PPEak AC positive peak spike volts measurements
[:FUNCtion]? Query selected measurement type
These commands are used to select the measurement type for the ACI and ACV functions. Refer
to paragraphs 2.4.1 (ACV) and 2.4.2 (ACI) for details on AC-TYPE.
PRINT #1, "output 16; :curr:ac:det aver; det?" ' Select average detector
PRINT #1, "enter 16" ' Get response from 2002
Program fragment
Parameters
Query
Description
Program fragment
3.18.12
Parameters
Query
Description
Program fragment

Table of Contents

Other manuals for Keithley 2002

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2002 and is the answer not in the manual?

Keithley 2002 Specifications

General IconGeneral
Resolution8.5 digits
DC Voltage Accuracy±0.0006%
Maximum DC Voltage1000 V
Maximum AC Voltage750 V
Maximum DC Current3 A
Maximum AC Current3 A
ConnectivityGPIB, RS-232
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Resistance Range10 Ω to 1 GΩ
Model2002
CategoryMultimeter
InterfaceGPIB, RS-232
AC Current Accuracy±0.1%
Frequency Range0.1Hz to 300kHz
DisplayVFD
TypeDigital
Power Supply100-240V AC, 50/60Hz
Current Range (DC)100 μA to 3 A
Current Accuracy (DC)±0.01%
Current Range (AC)100 μA to 3 A
Current Accuracy (AC)±0.1%
DC Current Range100 μA to 3 A
AC Current Range100 μA to 3 A

Related product manuals