EasyManuals Logo

Keithley 2002 User Manual

Keithley 2002
366 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #339 background imageLoading...
Page #339 background image
H
HP3458A Emulation Mode
H-1
Introduction
This appendix provides detailed information on using the
Model 2002 HP3458A emulation mode. The emulation
mode gives the Model 2002 a high degree of compatibility
with the Hewlett-Packard HP3458A Digital Multimeter,
allowing the instrument to be substituted for an HP3458A in
an existing measurement system with little or no program
modifications.
Firmware revision level
The HP3458A emulation mode is available in Model 2002s
equipped with revision A05 and later main processor firm-
ware. The firmware revision level is displayed during the
power-up cycle and may be displayed at any time by using
the GENERAL/ SERIAL # selection in the main MENU.
Enabling the HP3458A emulation mode
The HP3458A emulation mode may be enabled from the
front panel by using the LANGUAGE selection in the GPIB
menu as follows:
1. Press the MENU key to display the main menu.
2. Select GPIB, then press ENTER.
3. Select LANGUAGE, then press ENTER.
4. Choose HP3458-EMULATION, then press ENTER.
5. Enable or disable the emulation mode as desired, then
press ENTER to complete your selection.
6. Press EXIT as required to return to normal display.
Unsupported commands
Table H-1 summarizes commands not supported by the
Model 2002 HP3458A emulation mode.

Table of Contents

Other manuals for Keithley 2002

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2002 and is the answer not in the manual?

Keithley 2002 Specifications

General IconGeneral
Resolution8.5 digits
DC Voltage Accuracy±0.0006%
Maximum DC Voltage1000 V
Maximum AC Voltage750 V
Maximum DC Current3 A
Maximum AC Current3 A
ConnectivityGPIB, RS-232
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Resistance Range10 Ω to 1 GΩ
Model2002
CategoryMultimeter
InterfaceGPIB, RS-232
AC Current Accuracy±0.1%
Frequency Range0.1Hz to 300kHz
DisplayVFD
TypeDigital
Power Supply100-240V AC, 50/60Hz
Current Range (DC)100 μA to 3 A
Current Accuracy (DC)±0.01%
Current Range (AC)100 μA to 3 A
Current Accuracy (AC)±0.1%
DC Current Range100 μA to 3 A
AC Current Range100 μA to 3 A

Related product manuals