Front Panel Operation
2-87
4. If you wish to use the current default function for all
channels, select DEFAULT, then press ENTER.
5. If you intend to program a separate function for each
channel, choose CHOOSE-FUNCTIONS, then press
ENTER. The instrument will prompt you for a channel
number, as in the following typical display:
SELECT CHAN=01 (DCV)
6. Use the cursor and range keys to select a channel, then
press ENTER. Select the desired function, then press
ENTER. Repeat the procedure for all channels to be
defined.
7. Press EXIT as necessary to return to normal display.
Step 4: Enable external scanning
1. From normal display, press CONFIG-SCAN. The in-
strument will display the following:
SCAN OPERATION
INTERNAL EXTERNAL RATIO DELTA
2. Select EXTERNAL, then press ENTER.
Step 5: Start scan
Press SCAN for instructions to set up the external scanner,
program the scan count, scan timer, and data storage. Then
press ENTER to start scanning. When the scan has complet-
ed you can recall stored data and/or scan again. Press EXIT
to interrupt data storage and disable scanning.
2.12 Menu
The main menu accesses the various instrument operations
for which there are no dedicated keys, such as setup storage,
IEEE-488 setup, calibration, self-test, and limits. The main
menu structure is summarized in Table 2-41.
The top level of the main menu is displayed by placing the
instrument in the reading display state, and then pressing the
MENU key.
Some general rules to navigate the menu levels are given in
paragraph 2.3.4.

Table of Contents

Other manuals for Keithley 2002

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2002 and is the answer not in the manual?

Keithley 2002 Specifications

General IconGeneral
Resolution8.5 digits
DC Voltage Accuracy±0.0006%
Maximum DC Voltage1000 V
Maximum AC Voltage750 V
Maximum DC Current3 A
Maximum AC Current3 A
ConnectivityGPIB, RS-232
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Resistance Range10 Ω to 1 GΩ
Model2002
CategoryMultimeter
InterfaceGPIB, RS-232
AC Current Accuracy±0.1%
Frequency Range0.1Hz to 300kHz
DisplayVFD
TypeDigital
Power Supply100-240V AC, 50/60Hz
Current Range (DC)100 μA to 3 A
Current Accuracy (DC)±0.01%
Current Range (AC)100 μA to 3 A
Current Accuracy (AC)±0.1%
DC Current Range100 μA to 3 A
AC Current Range100 μA to 3 A

Related product manuals