IEEE-488 Reference
3-85
3.16 :OUTPut subsystem
The OUTPut subsystem is used to set polarities for the digital output port. Commands in this
subsystem are summarized in Table 3-8.
:LSENse <name>
:OUTPut:TTL[1]:LSENse <name> Set polarity of line #1
:OUTPut:TTL2:LSENse <name> Set polarity of line #2
:OUTPut:TTL3:LSENse <name> Set polarity of line #3
:OUTPut:TTL4:LSENse <name> Set polarity of line #4
<name> = AHIGh Set polarity of specified line to active high.
ALOW Set polarity of specified line to active low.
:LSENse? Query polarity of specified line
These commands are used to set the polarity of the digital output lines. When set for active high
(AHIGh) polarity, the specified output line is true (ON) when the output level is high. The out-
put line is false (OFF) when the output level is low. When set for active low (ALOW) polarity,
the output line is true (ON) when the output level is low. The output line is false (OFF) when
the output level is high.
The logic levels (true or false) of the digital output lines are set from the SOURce Subsystem
(see paragraph 3.19).
PRINT #1, "output 16; :outp:ttl:lsen alow; lsen?" ' Set line #1 to ALOW
PRINT #1, "enter 16" ' Get response message from
2002
Parameters
Query
Description
Program fragment

Table of Contents

Other manuals for Keithley 2002

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2002 and is the answer not in the manual?

Keithley 2002 Specifications

General IconGeneral
Resolution8.5 digits
DC Voltage Accuracy±0.0006%
Maximum DC Voltage1000 V
Maximum AC Voltage750 V
Maximum DC Current3 A
Maximum AC Current3 A
ConnectivityGPIB, RS-232
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Resistance Range10 Ω to 1 GΩ
Model2002
CategoryMultimeter
InterfaceGPIB, RS-232
AC Current Accuracy±0.1%
Frequency Range0.1Hz to 300kHz
DisplayVFD
TypeDigital
Power Supply100-240V AC, 50/60Hz
Current Range (DC)100 μA to 3 A
Current Accuracy (DC)±0.01%
Current Range (AC)100 μA to 3 A
Current Accuracy (AC)±0.1%
DC Current Range100 μA to 3 A
AC Current Range100 μA to 3 A

Related product manuals