IEEE-488 Reference
3-123
Figure 3-27
Measurement Event Enable Register
RAV
B15 - B12 B10 B9 B8 B7 B6 B5 B4 B3 B2 B1 B0
(2 )
9
(2 )
8
(2 )
7
(2 )
5
(2 )
4
(2 )
3
(2 )
2
(2 )
1
(2 )
0
32
0/1
Bit Position
Event
Decimal Weighting
Value
Value : 1 = Enable Measurement Event
0 = Disable (Mask) Measurement Event
LL1
2
0/1
Events : BPT = Buffer Pretriggered
BFL = Buffer Full
BHF = Buffer Half Full
BAV = Buffer Available
RAV = Reading Available
HL2 = High Limit 2
LL2 = Low Limit 2
HL1 = High Limit 1
LL1 = Low Limit 1
ROF = Reading Overflow
BFL HL2 HL1 ROF
512 16 4 1
0/1 0/1 0/1 0/1
BHF BAV LL2
256 128 8
0/1 0/1 0/1
BPT
B11
(2 )
11
2048
0/1
Figure 3-28
Questionable Event Enable Register
Bit Position
Event
Decimal Weighting
Value
Value : 1 = Enable Questionable Event
0 = Disable (Mask) Questionable Event
Events : Warn = Command Warning
Cal = Calibration Summary
Temp = Temperature Summary
Cal
B13 - B9 B8B15 B14
(2 )
8
256
(2 )
14
16384
0/10/1
Warn
B7 - B5
0
Temp
B4
(2 )
4
16
0/1
B3 - B0