IEEE-488 Reference
3-103
The number of readings that are averaged by the filter is set with the :AVERage:COUNt com-
mand. The :AVERage[:STATe] command is used to enable or disable the filter. Changing the
filter type disables auto filter.
PRINT #1, "output 16; :volt:dc:aver:tcon rep; tcon?" ' Select repeating
filter
PRINT #1, "enter 16" ' Get response from 2002
:COUNt <n>
[:SENSe[1]]:CURRent:AC:AVERage:COUNt <n> Specify filter count for ACI
[:SENSe[1]]:CURRent:DC:AVERage:COUNt <n> Specify filter count for DCI
[:SENSe[1]]:VOLTage:AC:AVERage:COUNt <n> Specify filter count for ACV
[:SENSe[1]]:VOLTage:DC:AVERage:COUNt <n> Specify filter count for DCV
[:SENSe[1]]:RESistance:AVERage:COUNt <n> Specify filter count for 2
[:SENSe[1]]:FRESistance:AVERage:COUNt <n> Specify filter count for 4
[:SENSe[1]]:TEMPerature:AVERage:COUNt <n> Specify filter count for TEMP
<n> = 1 to 100 Specify filter count
DEFault 10
MINimum 1
MAXimum 100
:COUNt? Query filter count
:COUNt? DEFault Query the *RST default filter count
:COUNt? MINimum Query the lowest allowable filter count
:COUNt? MAXimum Query the largest allowable filter count
These commands are used to specify the filter count. In general, the filter count is the number
of readings that are acquired and stored in the filter buffer for the averaging calculation. The
larger the filter count, the more filtering that is performed.
PRINT #1, "output 16; :volt:dc:aver:coun 20; coun?" ' Set filter count
PRINT #1, "enter 16" ' Get response from 2002
[:STATe] <b>
[:SENSe[1]]:CURRent:AC:AVERage[:STATe] <b> Control filter for ACI
[:SENSe[1]]:CURRent:DC:AVERage[:STATe] <b> Control filter for DCI
[:SENSe[1]]:VOLTage:AC:AVERage[:STATe] <b> Control filter for ACV
[:SENSe[1]]:VOLTage:DC:AVERage[:STATe] <b> Control filter for DCV
[:SENSe[1]]:RESistance:AVERage[:STATe] <b> Control filter for 2
[:SENSe[1]]:FRESistance:AVERage[:STATe] <b> Control filter for 4
[:SENSe[1]]:TEMPerature:AVERage[:STATe] <b> Control filter for TEMP
<b> = 0 or OFF Disable the digital filter
1 or ON Enable the digital filter
[:STATe]? Query state of digital filter
These commands are used to enable or disable the digital averaging filter for the specified func-
tion. When enabled, readings will be filtered according to how the filter is configured.
PRINT #1, "output 16; :volt:dc:aver on; aver?" ' Enable filter
PRINT #1, "enter 16" ' Get response from 2002
Program fragment
Parameters
Query
Description
Program fragment
Parameters
Query
Description
Program fragment

Table of Contents

Other manuals for Keithley 2002

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2002 and is the answer not in the manual?

Keithley 2002 Specifications

General IconGeneral
Resolution8.5 digits
DC Voltage Accuracy±0.0006%
Maximum DC Voltage1000 V
Maximum AC Voltage750 V
Maximum DC Current3 A
Maximum AC Current3 A
ConnectivityGPIB, RS-232
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Resistance Range10 Ω to 1 GΩ
Model2002
CategoryMultimeter
InterfaceGPIB, RS-232
AC Current Accuracy±0.1%
Frequency Range0.1Hz to 300kHz
DisplayVFD
TypeDigital
Power Supply100-240V AC, 50/60Hz
Current Range (DC)100 μA to 3 A
Current Accuracy (DC)±0.01%
Current Range (AC)100 μA to 3 A
Current Accuracy (AC)±0.1%
DC Current Range100 μA to 3 A
AC Current Range100 μA to 3 A

Related product manuals