IEEE-488 Reference
3-101
:DIGits <n>
[:SENSe[1]]:CURRent:AC:DIGits <n> Specify resolution for ACI
[:SENSe[1]]:CURRent:DC:DIGits <n> Specify resolution for DCI
[:SENSe[1]]:VOLTage:AC:DIGits <n> Specify resolution for ACV
[:SENSe[1]]:VOLTage:DC:DIGits <n> Specify resolution for DCV
[:SENSe[1]]:RESistance:DIGits <n> Specify resolution for 2
[:SENSe[1]]:FRESistance:DIGits <n> Specify resolution for 4
[:SENSe[1]]:FREQuency:DIGits <n> Specify resolution for FREQ
[:SENSe[1]]:TEMPerature:DIGits <n> Specify resolution for TEMP
<n> = 4 3½ digits for DCI, ACI, ACV, DCV, 2, 4
3½ digits (1° resolution) for TEMP
4 digits for FREQ
54½ digits for DCI, ACI, ACV, DCV, 2, 4
4½ digits (0.1° resolution) for TEMP
5 digits for FREQ
65½ digits for DCI, ACI, ACV, DCV, 2, 4
5½ digits (0.01° resolution) for TEMP
76½ digits for DCI, ACI, ACV, DCV, 2, 4
6½ digits (0.001° resolution) for TEMP
87½ digits for DCI, ACI, ACV, DCV, 2, 4
98½ digits for DCI, ACI, ACV, DCV, 2, 4
DEFault 7½ digits for DCV, 2 and 4
6½ digits for DCI and ACI
5½ digits for ACV
5½ digits (0.01° resolution) for TEMP
5 digits for FREQ
MINimum 3½ digits for DCI, ACI, ACV, DCV, 2, 4
3½ digits (1° resolution) for TEMP
4 digits for FREQ
MAXimum 8½ digits for DCI, ACI, ACV, DCV, 2, 4
6½ digits (0.001° resolution) for TEMP
5 digits for FREQ
:DIGits? Query selected resolution
:DIGits? DEFault Query *RST default resolution
:DIGits? MINimum Query minimum allowable resolution
:DIGits? MAXimum Query maximum allowable resolution
These commands are used to select display resolution for the specified measurement function.
Even though the parameters for this command are expressed as integers (4 to 9), you can specify
resolution using real numbers. For example, to select 3½ digit resolution let <n> = 3.5, for 4½d
let <n> = 4.5, and so on. Internally, the instrument rounds the entered parameter value to the
nearest integer.
PRINT #1, "output 16; :volt:dc:dig 4.5; dig?" ' Select 4½d DCV resolution
PRINT #1, "enter 16" ' Get response from 2002
3.18.10
Parameters
Query
Description
Program fragment

Table of Contents

Other manuals for Keithley 2002

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2002 and is the answer not in the manual?

Keithley 2002 Specifications

General IconGeneral
Resolution8.5 digits
DC Voltage Accuracy±0.0006%
Maximum DC Voltage1000 V
Maximum AC Voltage750 V
Maximum DC Current3 A
Maximum AC Current3 A
ConnectivityGPIB, RS-232
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Resistance Range10 Ω to 1 GΩ
Model2002
CategoryMultimeter
InterfaceGPIB, RS-232
AC Current Accuracy±0.1%
Frequency Range0.1Hz to 300kHz
DisplayVFD
TypeDigital
Power Supply100-240V AC, 50/60Hz
Current Range (DC)100 μA to 3 A
Current Accuracy (DC)±0.01%
Current Range (AC)100 μA to 3 A
Current Accuracy (AC)±0.1%
DC Current Range100 μA to 3 A
AC Current Range100 μA to 3 A