IEEE-488 Reference
3-100
:STATe <b>
[:SENSe[1]]:CURRent:AC:REFerence:STATe <b> Control reference for ACI
[:SENSe[1]]:CURRent:DC:REFerence:STATe <b> Control reference for DCI
[:SENSe[1]]:VOLTage:AC:REFerence:STATe <b> Control reference for ACV
[:SENSe[1]]:VOLTage:DC:REFerence:STATe <b> Control reference for DCV
[:SENSe[1]]:RESistance:REFerence:STATe <b> Control reference for 2
[:SENSe[1]]:FRESistance:REFerence:STATe <b> Control reference for 4
[:SENSe[1]]:FREQuency:REFerence:STATe <b> Control reference for FREQ
[:SENSe[1]]:TEMPerature:REFerence:STATe <b> Control reference for TEMP
<b> = 1 or ON Enable reference
0 or OFF Disable reference
:STATe? Query state of reference
These commands are used to enable or disable Reference for the specified function. When
enabled, the displayed reading will include the programmed reference value (see :REFerence
<n> and :ACQuire). When disabled, the displayed reading will not include the reference value.
PRINT #1, "output 16; :curr:ac:ref:stat on; stat?" ' Enable reference
PRINT #1, "enter 16" ' Get response from 2002
:ACQuire
[:SENSe[1]]:CURRent:AC:REFerence:ACQuire Acquire Reference for ACI
[:SENSe[1]]:CURRent:DC:REFerence:ACQuire Acquire Reference for DCI
[:SENSe[1]]:VOLTage:AC:REFerence:ACQuire Acquire Reference for ACV
[:SENSe[1]]:VOLTage:DC:REFerence:ACQuire Acquire Reference for DCV
[:SENSe[1]]:RESistance:REFerence:ACQuire Acquire Reference for 2
[:SENSe[1]]:FRESistance:REFerence:ACQuire Acquire Reference for 4
[:SENSe[1]]:FREQuency:REFerence:ACQuire Acquire Reference for FREQ
[:SENSe[1]]:TEMPerature:REFerence:ACQuire Acquire Reference for TEMP
When one of these commands is sent, the measured input signal is acquired and established as
the reference value. This command is typically used to zero the display. For example, if the
instrument is displaying a 2mA offset, sending this command and enabling Reference (see
:STATe) zeroes the display.
This command is functional only if the instrument is on the specified measurement function.
Sending this command while in any other function causes an error. Also, if the latest reading is
overflowed (“OFLO”) or a reading has not been triggered (“----”), an error occurs when this
command is sent.
The :ACQuire command is coupled to the :REFerence <n> command. See the description for
:REFerence for details.
PRINT #1, "output 16; :func 'volt:dc'" ' Select DCV function
SLEEP 1 ' Allow reading to settle
PRINT #1, "output 16; :volt:dc:ref:acq; stat on" ' Acquire and enable refer-
ence
PRINT #1, "output 16; :volt:dc:ref?" ' Query reference value
PRINT #1, "enter 16" ' Get response from 2002
Parameters
Query
Description
Program fragment
Description
Program fragment

Table of Contents

Other manuals for Keithley 2002

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2002 and is the answer not in the manual?

Keithley 2002 Specifications

General IconGeneral
Resolution8.5 digits
DC Voltage Accuracy±0.0006%
Maximum DC Voltage1000 V
Maximum AC Voltage750 V
Maximum DC Current3 A
Maximum AC Current3 A
ConnectivityGPIB, RS-232
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Resistance Range10 Ω to 1 GΩ
Model2002
CategoryMultimeter
InterfaceGPIB, RS-232
AC Current Accuracy±0.1%
Frequency Range0.1Hz to 300kHz
DisplayVFD
TypeDigital
Power Supply100-240V AC, 50/60Hz
Current Range (DC)100 μA to 3 A
Current Accuracy (DC)±0.01%
Current Range (AC)100 μA to 3 A
Current Accuracy (AC)±0.1%
DC Current Range100 μA to 3 A
AC Current Range100 μA to 3 A