EasyManuals Logo

Keithley 2002 User Manual

Keithley 2002
366 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #243 background imageLoading...
Page #243 background image
IEEE-488 Reference
3-120
Trigger Event Register:
Bit B0 —
Not used.
Bit B1, Sequence 1 (Seq1) —
Set bit indicates that the instrument is in the trigger layer (PTR),
or that the instrument has exited from the trigger layer (NTR).
Bits B2 through B15 —
Not used.
Arm Event Register:
Bit B0 —
Not used.
Figure 3-23
Operation Event Register
Idle Arm Trig
B15 - B11 B10 B9 B8 B7 B6 B5 B4 B3 B2 B1 B0
(2 )
10
1024
(2 )
6
(2 )
5
(2 )
0
64 32
0/1 0/1 0/1
Bit Position
Event
Decimal Weighting
Value
Value : 1 = Operation Event Set
0 = Operation Event Cleared
Events : Idle = Idle state of the 2002
Arm = Waiting for Arm
Trig = Waiting for Trigger
Meas = Measuring
Cal = Calibrating
Cal
1
0/1
Meas
(2 )
4
16
0/1
Figure 3-24
Trigger Event Register
B15 - B2
B1 B0
(2 )
1
2
0/1
Bit Position
Event
Decimal Weighting
Value
Value : 1 = Event Bit Set
0 = Event Bit Cleared
Event : Seq1 = Sequence 1
Seq1

Table of Contents

Other manuals for Keithley 2002

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2002 and is the answer not in the manual?

Keithley 2002 Specifications

General IconGeneral
Resolution8.5 digits
DC Voltage Accuracy±0.0006%
Maximum DC Voltage1000 V
Maximum AC Voltage750 V
Maximum DC Current3 A
Maximum AC Current3 A
ConnectivityGPIB, RS-232
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Resistance Range10 Ω to 1 GΩ
Model2002
CategoryMultimeter
InterfaceGPIB, RS-232
AC Current Accuracy±0.1%
Frequency Range0.1Hz to 300kHz
DisplayVFD
TypeDigital
Power Supply100-240V AC, 50/60Hz
Current Range (DC)100 μA to 3 A
Current Accuracy (DC)±0.01%
Current Range (AC)100 μA to 3 A
Current Accuracy (AC)±0.1%
DC Current Range100 μA to 3 A
AC Current Range100 μA to 3 A

Related product manuals