EasyManuals Logo

Keithley 2002 User Manual

Keithley 2002
366 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #191 background imageLoading...
Page #191 background image
IEEE-488 Reference
3-68
This command is used to specify the “a1” factor for the polynomial calculation.
PRINT #1, "output 16; :calc:kmat:ma1f 4; ma1f?" ' Set "a1" and query.
PRINT #1, "enter 16" ' Get response from 2002
:MA2Factor <NRf>
:CALCulate[1]:KMATh:MA2Factor <NRf> Specify “a2” factor
<NRf> = -9.999e20 to +9.999e20
:MA2Factor Query “a2” factor
This command is used to specify the “a2” factor for the polynomial calculation.
PRINT #1, "output 16; :calc:kmat:ma2f 6; ma0f?" ' Set "a2" and query.
PRINT #1, "enter 16" ' Get response from 2002
:PERCent <NRf>
:CALCulate[1]:KMATh:PERCent <NRf> Specify Percent target value
<NRf> = -9.999999e35 to +9.999999e35 Specify target value for Percent calculation.
:PERCent? Query Percent target value
This command is used to specify the target value for the Percent calculation.
PRINT #1, "output 16; :calc:kmat:perc 10; perc?" ' Set target to 10% and
query
PRINT #1, "enter 16" ' Get response message from
2002
:STATe <b>
:CALCulate[1]:STATe <b> Control CALC 1
<b> = 0 or OFF Disable CALC 1 calculation
1 or ON Enable CALC 1 calculation
:STATe? Query state (on or off) of CALC 1
This command is used to enable or disable the CALC 1 calculation. When enabled, each instru-
ment reading will reflect the selected calculation (see :FORMat).
PRINT #1, "output 16; :calc:stat on;stat?" ' Enable CALC1 and query
PRINT #1, "enter 16" ' Get response message from 2002
:DATA commands
[:LATest]?
:CALCulate[1]:DATA[:LATest]? Read CALC 1 result
This query command is used to read the result of the CALC 1 calculation. If CALC 1 is disabled
or NONE is selected, the “raw” reading will instead be read.
Description
Program fragment
Parameters
Query
Description
Program fragment
Parameters
Query
Description
Program fragment
Parameters
Query
Description
Program fragment
Description

Table of Contents

Other manuals for Keithley 2002

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2002 and is the answer not in the manual?

Keithley 2002 Specifications

General IconGeneral
Resolution8.5 digits
DC Voltage Accuracy±0.0006%
Maximum DC Voltage1000 V
Maximum AC Voltage750 V
Maximum DC Current3 A
Maximum AC Current3 A
ConnectivityGPIB, RS-232
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Resistance Range10 Ω to 1 GΩ
Model2002
CategoryMultimeter
InterfaceGPIB, RS-232
AC Current Accuracy±0.1%
Frequency Range0.1Hz to 300kHz
DisplayVFD
TypeDigital
Power Supply100-240V AC, 50/60Hz
Current Range (DC)100 μA to 3 A
Current Accuracy (DC)±0.01%
Current Range (AC)100 μA to 3 A
Current Accuracy (AC)±0.1%
DC Current Range100 μA to 3 A
AC Current Range100 μA to 3 A