EasyManuals Logo

Keithley 2002 User Manual

Keithley 2002
366 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #200 background imageLoading...
Page #200 background image
IEEE-488 Reference
3-77
PRINT #1, "output 16; :disp:text:stat on; stat?" ' Enable and query message
mode
PRINT #1, "enter 16" ' Get response message from
2002
:DATA?
:DISPlay[:WINDow[1]]:DATA? Read top display
:DISPlay:WINDow2:DATA? Read bottom display
These query commands are used to read what is currently being displayed on the top and bottom
displays. After sending one of these commands and addressing the Model 2002 to talk, the dis-
played data (message or reading) will be sent to the computer.
PRINT #1, "output 16; :disp:data?" ' Query displayed data
PRINT #1, "enter 16" ' Get response message from 2002
:CNDisplay
:DISPlay:CNDisplay Clear (cancel) NEXT display
This action command is used to clear the bottom display of NEXT (or PREV) messages and
cancel the operations associated with them. This command has no effect on any other message
types.
This example assumes that a NEXT operation is currently being displayed.
PRINT #1, "output 16; :cnd" ' Clear NEXT display
:SMESsage <b>
:DISPlay:SMESsage <b> Control status message
<b> = 0 or OFF Disable status messages
= 1 or ON Enable status messages
:SMESsage? Query state of status message mode
This command is used to enable/disable the status message mode. The status message mode is
a diagnostic tool that provides real-time messages that relate to the current operating state of the
instrument. Table 2-4 lists the status messages.
PRINT #1, "output 16; :disp:smes on; smes?" ' Enable and query status message
mode
PRINT #1, "enter 16" ' Get response message from 2002
:ENABle <b>
:DISPlay:ENABle <b> Control display circuitry
<b> = 0 or OFF Disable display circuitry
= 1 or ON Enable display circuitry
:ENABle? Query state of display
Program fragment
Description
Program fragment
Description
Program fragment
Parameters
Query
Description
Program fragment
Parameters
Query

Table of Contents

Other manuals for Keithley 2002

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2002 and is the answer not in the manual?

Keithley 2002 Specifications

General IconGeneral
Resolution8.5 digits
DC Voltage Accuracy±0.0006%
Maximum DC Voltage1000 V
Maximum AC Voltage750 V
Maximum DC Current3 A
Maximum AC Current3 A
ConnectivityGPIB, RS-232
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Resistance Range10 Ω to 1 GΩ
Model2002
CategoryMultimeter
InterfaceGPIB, RS-232
AC Current Accuracy±0.1%
Frequency Range0.1Hz to 300kHz
DisplayVFD
TypeDigital
Power Supply100-240V AC, 50/60Hz
Current Range (DC)100 μA to 3 A
Current Accuracy (DC)±0.01%
Current Range (AC)100 μA to 3 A
Current Accuracy (AC)±0.1%
DC Current Range100 μA to 3 A
AC Current Range100 μA to 3 A

Related product manuals