'PRINT #1, "output 16; :rout:scan:rat:mch (@ 2); mch?" ' Select and query
reference channel
PRINT #1, "enter 16" ' Get response mes-
sage from 2002
:LSELect <name>
:ROUTe:SCAN:LSELect <name> Perform specified scan operation
<name> = INTernal Enable scan for internal scanner card
EXTernal Enable scan for external scanner card
RATio Perform ratio calculation
DELTa Perform delta calculation
NONE Disable all scan operations
:LSELect? Query scan operation
This command is used to select and perform the desired scan operation. When INTernal is se-
lected, the Model 2002 scans the channels of the internal switching card according to how the
scan is configured (see :ROUTe:SCAN[:INTernal]). EXTernal is used to measure channels that
are controlled by an external switch system. When EXTernal is selected, the Model 2002 scans
the external scan list (see :SCAN:EXTernal).
With RATio or DELTa selected, the Model 2002 performs the appropriate math calculation us-
ing the specified reference channel and measure channel (see :RCHannel and :MCHannel).
The NONE selection disables all operations associated with a scan.
PRINT #1, "output 16; :syst:pres" ' Return 2002 to defaults
PRINT #1, "output 16; :rout:scan:lse1 int" ' Scan internal channels
Program fragment
Parameters
Query
Description
Program fragment

Table of Contents

Other manuals for Keithley 2002

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2002 and is the answer not in the manual?

Keithley 2002 Specifications

General IconGeneral
Resolution8.5 digits
DC Voltage Accuracy±0.0006%
Maximum DC Voltage1000 V
Maximum AC Voltage750 V
Maximum DC Current3 A
Maximum AC Current3 A
ConnectivityGPIB, RS-232
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Resistance Range10 Ω to 1 GΩ
Model2002
CategoryMultimeter
InterfaceGPIB, RS-232
AC Current Accuracy±0.1%
Frequency Range0.1Hz to 300kHz
DisplayVFD
TypeDigital
Power Supply100-240V AC, 50/60Hz
Current Range (DC)100 μA to 3 A
Current Accuracy (DC)±0.01%
Current Range (AC)100 μA to 3 A
Current Accuracy (AC)±0.1%
DC Current Range100 μA to 3 A
AC Current Range100 μA to 3 A