IEEE-488 Reference
3-136
Note: To disable all messages from entering the Error Queue, send the following command:
:stat:que:enab ()
PRINT #1, "output 16; :stat:que:enab (0:999); enab?" ' Enable all messages
and query
PRINT #1, "enter 16" ' Get response message
from 2002
DISable <list>
:STATus:QUEue:DISable <list> Disable messages for Error Queue
<list> = (numlist)
where; numlist is a specified list of messages that you wish to disable for the Error Queue.
:DISable? Query list of disabled messages
On power-up, all error messages are enabled and will go into the Error Queue as they occur. Sta-
tus messages are not enabled and will not go into the queue. This command is used to specify
which messages you want disabled. Disabled messages are prevented from going into the Error
Queue.
Messages are specified by numbers (see Table 2-4). See :QUEue:ENABle for examples to ex-
press a numlist.
PRINT #1, "output 16; :stat:que:dis (-140:-150); dis?" ' Disable messages
and query
PRINT #1, "enter 16" ' Get response mes-
sage from 2002
:CLEar
:STATus:QUEue:CLEar Clear Error Queue
This action command is used to clear the Error Queue of messages.
PRINT #1, "output 16; :stat:que:cle" ' Clear Error Queue
Program fragment
Parameters
Query
Description
Program fragment
Description
Program fragment

Table of Contents

Other manuals for Keithley 2002

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2002 and is the answer not in the manual?

Keithley 2002 Specifications

General IconGeneral
Resolution8.5 digits
DC Voltage Accuracy±0.0006%
Maximum DC Voltage1000 V
Maximum AC Voltage750 V
Maximum DC Current3 A
Maximum AC Current3 A
ConnectivityGPIB, RS-232
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Resistance Range10 Ω to 1 GΩ
Model2002
CategoryMultimeter
InterfaceGPIB, RS-232
AC Current Accuracy±0.1%
Frequency Range0.1Hz to 300kHz
DisplayVFD
TypeDigital
Power Supply100-240V AC, 50/60Hz
Current Range (DC)100 μA to 3 A
Current Accuracy (DC)±0.01%
Current Range (AC)100 μA to 3 A
Current Accuracy (AC)±0.1%
DC Current Range100 μA to 3 A
AC Current Range100 μA to 3 A