EasyManuals Logo

Keithley 2002 User Manual

Keithley 2002
366 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #281 background imageLoading...
Page #281 background image
IEEE-488 Reference
3-158
This command is used to set the date for the RTCLock control source. Setting an invalid date
(i.e. February 29, 1994) for the clock event results in an error and the previous date is retained.
PRINT #1, "output 16; :arm:rtcl:date <1993>, <12>, <1>; date?" ' Set date to
December 1,
1993 and query
PRINT #1, "enter 16" ' Get response
message from
2002
:TIME <hr>, <min>, <sec>
:ARM[:SEQuence[1]][LAYer[]]:RTCLock:TIME <hr>, <min>, <sec> Set time for RTCLock control source
<hr> = 0 to 23 Specify hour (24 hour format)
<min> = 0 to 59 Specify minute
<sec> = 0.00 to 59.99 Specify second (round to 1/100 sec.)
:DATE? Query the time
This command is used to set the time for the RTCLock control source. Note that the time must
be set using the 24-hour format (i.e. hour 13 is 1pm). Setting an invalid time results in an error
and the previous time is retained.
:TIME? returns the time to nearest hundreth of a second.
PRINT #1, "output 16; :arm:rtcl:time <14>, <36>, <0>; time?" ' Set time to
2:36 pm and
query
PRINT #1, "enter 16" ' Get response
message from
2002
Description
Program fragment
Parameters
Query
Description
Program fragment

Table of Contents

Other manuals for Keithley 2002

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2002 and is the answer not in the manual?

Keithley 2002 Specifications

General IconGeneral
Resolution8.5 digits
DC Voltage Accuracy±0.0006%
Maximum DC Voltage1000 V
Maximum AC Voltage750 V
Maximum DC Current3 A
Maximum AC Current3 A
ConnectivityGPIB, RS-232
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Resistance Range10 Ω to 1 GΩ
Model2002
CategoryMultimeter
InterfaceGPIB, RS-232
AC Current Accuracy±0.1%
Frequency Range0.1Hz to 300kHz
DisplayVFD
TypeDigital
Power Supply100-240V AC, 50/60Hz
Current Range (DC)100 μA to 3 A
Current Accuracy (DC)±0.01%
Current Range (AC)100 μA to 3 A
Current Accuracy (AC)±0.1%
DC Current Range100 μA to 3 A
AC Current Range100 μA to 3 A

Related product manuals