HP3458A Emulation Mode
H-4
APER
Purpose
To program the A/D integration time.
Format
APER <aperture>
Parameter
<aperture>
= 166.67E-6s to 1s
Default
1.67E-4s
Query
APER?
Returns aperture in seconds.
Description
The
APER
command programs the Model 2002 A/D integration period. The unit will accept values
between 0s and 1s, but it rounds values to meet the Model 2002’s valid parameter range.
Example
APER 1E-3
Program 1ms aperture.
ARANGE
Purpose
To enable/disable the autorange mode.
Format
ARANGE <range>
Parameter
<range>
<range> Parameter Numeric Equivalent Description
OFF 0 Disable autorange.
ON 1 Enable autorange.
ONCE 2 Autorange next reading.
Default
OFF
Query
ARANGE?
Returns autorange status.
Description
ARANGE
enables/disables autorange mode. The valid parameters are OFF, ON, and ONCE, which
correspond numerically to 0, 1, and 2 respectively.
Purpose
Autorange can also be controlled by the
FUNC
and
RANGE
commands. Specifying AUTO for the sec-
ond parameter of the
FUNC
command turns autoranging ON. Specifying AUTO for the first parameter
of the
RANGE
command turns autoranging ON.
Example
ARANGE ON
Enable autorange.
ARANGE 0
Disable autorange.
AZERO
Purpose
To enable/disable the autozero mode.
Format
AZERO <mode>
Parameter
<mode>

Table of Contents

Other manuals for Keithley 2002

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2002 and is the answer not in the manual?

Keithley 2002 Specifications

General IconGeneral
Resolution8.5 digits
DC Voltage Accuracy±0.0006%
Maximum DC Voltage1000 V
Maximum AC Voltage750 V
Maximum DC Current3 A
Maximum AC Current3 A
ConnectivityGPIB, RS-232
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Resistance Range10 Ω to 1 GΩ
Model2002
CategoryMultimeter
InterfaceGPIB, RS-232
AC Current Accuracy±0.1%
Frequency Range0.1Hz to 300kHz
DisplayVFD
TypeDigital
Power Supply100-240V AC, 50/60Hz
Current Range (DC)100 μA to 3 A
Current Accuracy (DC)±0.01%
Current Range (AC)100 μA to 3 A
Current Accuracy (AC)±0.1%
DC Current Range100 μA to 3 A
AC Current Range100 μA to 3 A

Related product manuals