HP3458A Emulation Mode
H-19
STB?
Purpose To request the status register value.
Format STB?
Response Decimal weight Bit number Status register condition
4 2 SRQ command executed
8 3 Power on
16 4 Ready for instructions
32 5 Error (as set by EMASK)
64 6 Service requested (SRQ)
128 7 Data available
Description The STB? query returns a number representing the weighted sum of all set bits in the status register.
The instrument may be programmed to generate an SRQ for specific status register conditions with
the RQS command.
Note The STB? query will always show bit 4 as being cleared since the query is being processed, but this
bit can be checked by using serial polling.
SWEEP
Purpose To program the interval between readings and the number of samples per trigger.
Format SWEEP <effective_interval>, <#_samples>
Parameter <effective_interval> 0.001 to 9999999.99
<#_samples> 1 to 99999
Default Interval: 0
#samples: 1024 (power-on=1)
Query SWEEP? Returns two responses separated by a comma: (1) the specified interval, and (2) the
specified number of samples.
Description The SWEEP command specifies the both the effective interval between readings (samples and the total
number of samples taken per trigger event.
Example SWEEP 0.1, 100 0.1 interval, 100 samples.
T or TRIG
Purpose To program the trigger event.
Format T <event>
TRIG <event>

Table of Contents

Other manuals for Keithley 2002

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2002 and is the answer not in the manual?

Keithley 2002 Specifications

General IconGeneral
Resolution8.5 digits
DC Voltage Accuracy±0.0006%
Maximum DC Voltage1000 V
Maximum AC Voltage750 V
Maximum DC Current3 A
Maximum AC Current3 A
ConnectivityGPIB, RS-232
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Resistance Range10 Ω to 1 GΩ
Model2002
CategoryMultimeter
InterfaceGPIB, RS-232
AC Current Accuracy±0.1%
Frequency Range0.1Hz to 300kHz
DisplayVFD
TypeDigital
Power Supply100-240V AC, 50/60Hz
Current Range (DC)100 μA to 3 A
Current Accuracy (DC)±0.01%
Current Range (AC)100 μA to 3 A
Current Accuracy (AC)±0.1%
DC Current Range100 μA to 3 A
AC Current Range100 μA to 3 A