HP3458A Emulation Mode
H-21
TEMP?
Purpose To request the internal temperature of the Model 2002.
Format TEMP?
Response Model 2002 internal temperature in °C
Description The TEMP? query returns the internal temperature of the Model 2002. The response is in standard
numeric format in degrees Celsius.
TERM
Purpose To request INPUTS switch status.
Format TERM <source>
Parameter <source> Not used
Query TERM? Returns status of INPUTS switch (1= Front; 2=Rear)
Description The TERM command is a no-op, and thus it is ignored. It is included for HP3458A compatibility. The
TERM? query returns the status of the INPUTS switch as indicated above.
TIMER
Purpose To define the timer sample interval.
Format TIMER <time>
Parameter <time> 0s to 6000s
Default 1s (power-on = 0s)
Query TIMER? Returns the currently specified time interval in seconds.
Description The TIMER command defines the time interval in seconds for the TIMER sample event in the NRDGS
command. The Model 2002 will accept time values from 0s to 6000s, but will modify values appro-
priately for compatibility.
Example TIMER 2 Define 2 second sample interval.

Table of Contents

Other manuals for Keithley 2002

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2002 and is the answer not in the manual?

Keithley 2002 Specifications

General IconGeneral
Resolution8.5 digits
DC Voltage Accuracy±0.0006%
Maximum DC Voltage1000 V
Maximum AC Voltage750 V
Maximum DC Current3 A
Maximum AC Current3 A
ConnectivityGPIB, RS-232
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Resistance Range10 Ω to 1 GΩ
Model2002
CategoryMultimeter
InterfaceGPIB, RS-232
AC Current Accuracy±0.1%
Frequency Range0.1Hz to 300kHz
DisplayVFD
TypeDigital
Power Supply100-240V AC, 50/60Hz
Current Range (DC)100 μA to 3 A
Current Accuracy (DC)±0.01%
Current Range (AC)100 μA to 3 A
Current Accuracy (AC)±0.1%
DC Current Range100 μA to 3 A
AC Current Range100 μA to 3 A

Related product manuals