Triple-Channel Power Supplies User's Manual Section 5:
2230G-900-01 Rev. A / June 2018 5-49
Example
STAT:QUES:INST:ISUM2:ENAB 12
Sets the third bit and fourth bit of the Questionable Instrument Summary Enable Register of channel 2
to 1, and sets all other bits to 0.
See also
None
STATus:QUEStionable:INSTrument:ISUMmary<X>[:EVENt]?
This command queries the Operation Instrument Summary Event Register of the status model for the specified
channel.
Group
Status
Syntax
STATus:QUEStionable:INSTrument:ISUMmary<X>[:EVENt]?
Arguments
<X> Channel number (1, 2, or 3)
Default value
0
Returns
<NR1> Range: 0 to 65535
Example
STAT:QUES:INST:ISUM1?
Queries the Questionable Instrument Summary Event Register of channel 1.
See also
None