5: SCPI command reference Series 2230G Triple-Channel Power Supplies
5-14 2230G-900-01 Rev. A / June 2018
Example
CAL:SAV
See also
None
CALibrate:SECure[:STATe]
This command enables or disables calibration mode.
Group
Calibration
Syntax
CALibrate:SECure[:STATe] <state> <password>
Arguments
<state> = <0|1> 0 enables the calibration mode, 1 disables calibration mode.
<password> = ”2230” The password is the model number of the power supply.
Default value
1
Returns
Not applicable
Example
CAL:SEC 0, "2230"
The calibration mode is enabled, and the password is 2230.
See also