Triple-Channel Power Supplies User's Manual Section 5:
2230G-900-01 Rev. A / June 2018 5-39
STATus subsystem
The STATus subsystem controls the status registers of the power supply.
STATus:OPERation:CONDition?
This command reads the Operation Condition Register of the status model.
Group
Status
Syntax
STATus:OPERation:CONDition?
Arguments
None
Default value
0
Returns
<NR1> Range: 0 to 255
Example
STAT:OPER:COND?
Returns the contents of the Operation Condition Register.
See also
None
STATus:OPERation:ENABle
This command sets or reads the Operation Event Enable Register of the status model.
Group
Status
Syntax
STATus:OPERation:ENABle <NR1>
STATus:OPERation:ENABle?
Arguments
<NR1> Range: 0 to 255
Default value
0