Keithley 2230G-30-3 - Measure[:Scalar][:Voltage][:DC]

Keithley 2230G-30-3
111 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Series 2230G
Triple-Channel Power Supplies User's Manual Section 5:
SCPI command reference
2230G-900-01 Rev. A / June 2018 5-25
Example
MEAS:POW? ALL
Makes a new power measurement on all channels.
See also
FETCh[:SCALar]:POWer[:DC]? (on page 5-22)
MEASure[:SCALar][:VOLTage][:DC]?
This command initiates and executes a new voltage measurement or queries the new measured voltage.
Group
Measurement
Syntax
MEASure[:SCALar][:VOLTage][:DC]? {CH1|CH2|CH3|ALL}
Arguments
CH1|CH2|CH3|ALL The channel or channels on which to make a new voltage measurement.
Default value
Not applicable
Returns
<NR2> The newly measured voltage with the unit of measure V.
Example
MEAS:VOLT? ALL
Makes a new voltage measurement on all channels.
See also
FETCh[:SCALar][:VOLTage][:DC]? (on page 5-23)