Triple-Channel Power Supplies User's Manual Section 5:
2230G-900-01 Rev. A / June 2018 5-15
CALibrate:STRing
This command writes or queries the calibration information of the instrument. The query version of this command
returns the saved calibration information for the instrument.
Group
Calibration
Syntax
CALibrate:STRing <string>
CALibrate:STRing?
Arguments
<string> The maximum length of the string is 22 bytes.
Default value
Not applicable
Returns
A string with a length of up to 22 bytes.
Example
CAL:STR "2017-03-28"
Sets the calibration information of the instrument to 2017-03-28.
See also
None
CALibrate:VOLTage[:DATA]
This command sets the actual output voltage value of the calibration point.
Group
Calibration
Syntax
CALibrate:VOLTage[:DATA] <NR2>
Arguments
<NR2> The voltage value of the calibration point.
Default value
Not applicable
Returns
Not applicable