EasyManuals Logo

Keithley 2230G-30-3 User Manual

Keithley 2230G-30-3
111 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #78 background imageLoading...
Page #78 background image
Section
5: SCPI command reference Series 2230G Triple-Channel Power Supplies
User's Manual
5-26 2230G-900-01 Rev. A / June 2018
SOURce subsystem
The commands in the SOURce subsystem are used to control the output of the power supply.
[SOURce:]APPLy
This command sets voltage and current levels on a specified channel with a single command message.
Group
Source
Syntax
[SOURce:]APPLy {CH1|CH2|CH3}, <NRf+>, <NRf+>
Arguments
CH1|CH2|CH3 The channel to apply the settings to
<NRF+> The voltage value to apply; valid values are MAX|MIN|DEF|UP|DOWN
<NRF+> The current value to apply; valid values are MAX|MIN|DEF|UP|DOWN
Default value
CH1, 1 V, 0.1 A
Returns
Not applicable
Example
APPL CH1, 15.0, 1.0
Sets the voltage and current levels on channel 1.
See also
INSTrument:NSELect (on page 5-20)
INSTrument[:SELect] (on page 5-21)
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude] (on page 5-28)
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude] (on page 5-35)

Table of Contents

Other manuals for Keithley 2230G-30-3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2230G-30-3 and is the answer not in the manual?

Keithley 2230G-30-3 Specifications

General IconGeneral
BrandKeithley
Model2230G-30-3
CategoryPower Supply
LanguageEnglish

Related product manuals