If the power supply does not turn on
To troubleshoot a power-on failure:
1. Verify that the AC power cord is firmly plugged into the power connector on the rear-panel of the
Series 2230G.
2. Make sure that the AC power source you plugged the power supply into is energized, and then
turn the Series 2230G on.
3. Make sure the voltage selector switch is set according to the local voltage (110 VAC or 220 VAC).
Change the voltage setting if it is incorrect.
4. Verify that the correct power-line fuse is installed and functioning. If the fuse is damaged, replace
it with a fuse appropriate for your model of power supply (see the following table).
Model number Specifications (220 VAC) Specifications (110 VAC)
2230G-30-3
3.15 A T250 V
6.30 A T250 V
2230G-30-6 5 A T250 V 10 A T250 V
2230G-60-3
5 A T250 V
10 A T250 V
5. If you need additional assistance, contact Keithley Instruments.
Self-test procedure
The Series 2230G runs a self-test automatically when it is turned on.
When the Series 2230G is turned on, the display shows the software version number (for example,
BOIS Ver. 1.XX). After approximately one second, the system starts the self-test and the display
shows Initializing….
If the self-test finishes with no errors, the display shows the set voltage value of all three channels on
the first line and the set current value of all three channels on the second line, in the format shown
below.
0.000V 0.000V 0.000V
SV0.000A SV0.000A SV0.000A
Self-test error messages
If an error occurs during the self-test, an error message is displayed. The following table lists the error
messages you might see.
Error message Meaning
EEPROM Test Failure
The EEPROM is damaged.
User Data Lost
The latest operation state of the power supply is lost.
Channel Initialization Failed
A channel failed to respond to inbound data.
Calibration Data Lost
Calibration data cannot be read.
Factory Calibration Data Lost
The factory calibration data in EEPROM is lost.

Table of Contents

Other manuals for Keithley 2230G-30-3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2230G-30-3 and is the answer not in the manual?

Keithley 2230G-30-3 Specifications

General IconGeneral
BrandKeithley
Model2230G-30-3
CategoryPower Supply
LanguageEnglish

Related product manuals